In hot/wet gas sampling systems, the sample must remain above its dew point from the probe through the sample line, pump and downstream analytical equipment. If the gas cools below its dew point, condensation may affect the pump, sample path and analyzer performance.
A heated pump head maintains the required temperature through the pump as part of a temperature-controlled gas sampling system.
Choose the Right Thermal Configuration
Heated Head
Use a Heated Head configuration when the gas sample must remain above its dew point through the pump in a standard hot/wet sampling application. Current product guidance identifies a typical application range of approximately 50-120°C.
Elevated Head
For higher-temperature applications, particularly where the installation environment or hazardous-area requirements do not allow the motor area to be heated, an Elevated Head configuration may be required. Current product guidance identifies a typical range of approximately 120-180°C.
Extended or Steam-Traced Head
For very wet or high-temperature sampling systems approaching 200°C, consult Air Dimensions about an Extended or steam-traced head configuration.